MAUDE Adverse Event Report: MICROVENTION, INC. HYDROFRAME 10 ADVANCED; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. HYDROFRAME 10 ADVANCED; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number 7110-0817-A2
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the coil was placed in the aneurysm and detached; however, during retrieval of the pusher wire post detachment, the physician noticed a strand of coil still connected to the wire.The majority of the coil detached and was implanted, but one strand / coil fiber remained attached to the pusher wire.Four (4) more coils were placed after, and the procedure was successfully completed.There was no reported injury or harm to the patient reported.
 
Additional Manufacturer Narrative
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was returned to the manufacturer for analysis and the investigation is currently underway.A supplemental report will be submitted when the investigation has been completed.
 
Additional Manufacturer Narrative
The investigation of the returned coil system found, the pusher severely stretched and broken at the distal section with lead wires exposed and broken.The implant was not returned for evaluation.However, the pusher's monofilament profile indicated, the implant was successfully detached from thermal activation, using a detachment controller.No strand of the implant coil was found attached to the pusher, as described in the reported event.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYDROFRAME 10 ADVANCED
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key15442846
Report Number2032493-2022-00369
Device Sequence Number5673443
Product Code HCG
UDI-Device Identifier00816777023735
UDI-Public(01)00816777023735(11)200109(17)241231(10)2001091WF
Combination Product (Y/N)Y
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K090357
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Model Number7110-0817-A2
Device Lot Number2001091WF
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2022
Supplement Date Received by Manufacturer10/19/2022
Initial Report FDA Received Date09/16/2022
Supplement Report FDA Received Date11/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-