DEPUY IRELAND - 9616671 ATUN PRESSFIT STR STEM14X160MM; KNEE STEMS Back to Search Results
Model Number 1513-14-160
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Swelling/ Edema (4577)
Date of Event 09/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient had implants removed and replaced due to suspected infection.Implants were placed on 2 dates per the records, (b)(6) 2019(first 5 items listed) and (b)(6) 2022.Patient experienced repeated effusions.Surgeon has sent multiple aspirations but no cultures were not positive.Cell counts and lab values were elevated, but not reaching a level to diagnose infection.During the procedure, 2 separate tissue samples were sent to pathology, and 3 and 4 wbc/hpf were found, again elevated but not definitive.Surgeon chose to removal all components and replace in a one stage fashion after a thorough debridement and irrigation.Antibiotic beads were placed in canals and surrounding soft tissues and vancomycin was added to gent cement.There was no surgical delay.Doi: (b)(6) 2019.Dor: (b)(6) 2022.Affected side: left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H10 additional narrative: added: d10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATUN PRESSFIT STR STEM14X160MM
Common Device Name
KNEE STEMS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15443147
Report Number1818910-2022-18136
Device Sequence Number5845162
Product Code JWH
UDI-Device Identifier10603295043645
UDI-Public10603295043645
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1513-14-160
Device Catalogue Number151314160
Device Lot NumberJ54C37
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2022
Supplement Date Received by Manufacturer09/22/2022
Initial Report FDA Received Date09/17/2022
Supplement Report FDA Received Date09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/11/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE REV LPS INSRT XSM 16MM; ATUN PRESSFIT STR STEM14X160MM; ATUN TIB SLV M/L 45MM FULL POR; ATUNE REV RP TIB BASE SZ 6 CEM; CONTROL CABLE W/CABLE SLEEVE; LPS DISTAL FEM COMP XSM LT; LPS FEM TO SLEEVE ADAPTER +0; LPS SEGMENTAL COMPONENT 25MM; LPS UNIV TIB HIN INS XSM 12MM; PFC*SIGMA/OV/DOME PAT 3PEG,41; UNIVERSAL FEM SLV FUL POR 46MM; UNIVERSAL STEM 75X20MM FLUTED; UNKNOWN BONE CEMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexFemale
Patient Weight115 KG
-
-