• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY TORIC EXTENDED VISION UV ABSORBING IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY TORIC EXTENDED VISION UV ABSORBING IOL; INTRAOCULAR LENS Back to Search Results
Model Number DAT315
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Disturbances (2140); Excessive Tear Production (2235); Irritability (2421); Eye Pain (4467); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and the documentation indicated the product met release criteria.The root cause has not been identified.There have been no other complaints reported in the lot.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-healthcare professional reported that following the cataract surgery with intraocular lens (iol) implant procedure the patient experienced blurry fuzzy edges.The patient cannot read a tv screen.Even large screen tvs have like the weather report even 8 to 10 foot away.Everything has a slight shadow just above it.Letters have a ghost shadow just above them.The biggest problem the patient facing is the dysphotopsia.All lights especially in the distance, greater than reading distance, have starbursts, glare, brilliance, sparkle, etc, hence can't read info from light.Additional information received and stated from the beginning everything is still blurry and out of focus.The edges of anything are not sharp and the contrast is poor.Eyes are very often sore, burning and stinging, sometime scratchy feeling.One evening after working in shop really straining to see, vision just went grey shadowy.There are two medical device reports associated with this event.This report is associated with the left eye.
 
Additional Manufacturer Narrative
Additional information provided in b.5 and h.6.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information received and stated that eye was tearing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ VIVITY TORIC EXTENDED VISION UV ABSORBING IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15443375
Report Number1119421-2022-01985
Device Sequence Number20229511
Product Code POE
UDI-Device Identifier00380652427135
UDI-Public00380652427135
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Model NumberDAT315
Device Lot Number15235248
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/22/2022
Supplement Date Received by Manufacturer09/19/2022
Initial Report FDA Received Date09/18/2022
Supplement Report FDA Received Date10/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-