• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI L.P. BURR, 6MM ROUND CUTTING 10-25 MM VAR EXPOSURE, USE WITH MDSSA AND MDASA; BUR, EAR, NOSE AND THROAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI L.P. BURR, 6MM ROUND CUTTING 10-25 MM VAR EXPOSURE, USE WITH MDSSA AND MDASA; BUR, EAR, NOSE AND THROAT Back to Search Results
Model Number MBUR6060FRCV
Medical Device Problem Codes Mechanical Problem (1384); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: (concomitant devices)-= short, angled handpiece mdasa sn:(b)(4), motor handpiece egmdhp200 sn:(b)(4) and 5mm drill egmbur5060frcv.The subject device has not yet been received for evaluation.Device is being shipped to the legal manufacturer for investigation.To date, the device has not yet been received.Note: photo of the subject device was provided.Communication with the customer via service department, rep (tm), the following information conveyed: patient was under anesthesia at the time of the event.According to the customer, delay in the procedure due to the event caused no harm to patient (prolongation of the procedure not considered by the user as a serious deterioration of a person's health ("clinically relevant") to which this medical device may have contributed.Procedure involved was a mastoidectomy.According to the reporter, "the impact of the drill did not cause any injury to the patient".Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Event or Problem Description
As reported , the 6mm drill deflects in the handpiece during use.The drill jumps a few millimeters to the side.The pressure with the drill on the material to be drilled was only very low.This happened several times happened during the surgery.Excessive heating was not detected.The short, angled handpiece mdasa sn:(b)(4) was used with the motor handpiece egmdhp200 sn:(b)(4).The user then used the 5mm drill egmbur5060frcv in the same procedure and had no problems with it.There was no patient harm, no user injury reported due to the event.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and the device evaluation.The device was returned to olympus for inspection, and the customer's complaint was not confirmed.The burr(mbur6060frcv) device was not returned in the original packaging.The burr appears to be unused with no presence of dents, markings or chips off the head of the burr.The locking grooves are in the correct placement on the burr and the tin coated tip is located correctly on the burr.Visually there is no failure present on the burr to cause the burr not to function as expected with the attachment and handpiece.There were no other abnormalities noted in the inspection.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the reported failure is a result of customer usage and assembly.The root cause has been determined to be customer assembly of the burr into the attachment and attachment to the handpiece.If the attachment is not locked into the handpiece properly then the burr (mbur6060frcv) cannot be inserted and locked into the attachment correctly.This will cause the burr(mbur6060frcv) to possibly dislodge out of the attachment and or "jump" while using the device.Proper assembly of the attachment is critical for the device to function properly with the handpiece.Olympus will continue to monitor field performance for this device.
 
Additional Manufacturer Narrative
This report is being submitted to correct the legal manufacturer¿s contact information and facility registration number.The facility registration number is (b)(4). .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BURR, 6MM ROUND CUTTING 10-25 MM VAR EXPOSURE, USE WITH MDSSA AND MDASA
Common Device Name
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
GYRUS ACMI L.P.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
GYRUS ACMI L.P.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
2925 appling rd
bartlett, TN 38133
5082077661
MDR Report Key15443427
Report Number1037007-2022-00017
Device Sequence Number14640147
Product Code EQJ
UDI-Device Identifier00821925036536
UDI-Public00821925036536
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT-EQJ
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMBUR6060FRCV
Device Lot NumberAV121761
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2022
Supplement Date Received by Manufacturer12/19/2022
06/18/2024
Initial Report FDA Received Date09/18/2022
Supplement Report FDA Received Date01/18/2023
06/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MDASA EGMDHP200 EGMBUR5060FRCV