| Model Number |
EIC7070-01 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an ent procedure, the electrode of the procise wand broke off, the broken parts were removed from the patient using tweezers.The procedure was successfully completed with non-significant delay using a back-up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).H10 h3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A visual inspection of the returned instrument shows no manufacturing abnormalities.The electrode is completely detached.Product was out of the original packaging.No packaging returned.A review of the customer provided image shows a used wand without packaging.The device was plugged into the controller and registered settings (7,3).The wand was not able to generate plasma and coagulation due to detached electrode.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states per case details, the broken piece was retrieved from the patient.The procedure was completed using a back-up device.No patient injuries or adverse consequences were reported.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.The complaint was confirmed and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include: (1) hitting the device tip against a hard surface.(2) using the device as a lever to enlarge surgical site.(3) mechanical displacement of tissue through applied force.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|