MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 BROVIAC CV CATHETER, SINGLE-LUMEN, 2.7F; CHRONIC CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 BROVIAC CV CATHETER, SINGLE-LUMEN, 2.7F; CHRONIC CATHETERS Back to Search Results
Model Number 0600040
Medical Device Problem Codes Fluid/Blood Leak (1250); Fracture (1260); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that sometimes post port placement, the line allegedly found to have a leak.It was further reported that catheter allegedly had hole.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.Expiry date: 08/2022.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: two bd posi-flush syringes, a extension set and one broviac s/l catheter were returned for evaluation.Gross, microscopic visual, functional testing were performed.The investigation is unconfirmed for the reported material hole issue as longitudinal split was identified on the catheter body.However, the investigation is confirmed for the reported fluid leak and identified split in the catheter issues as a longitudinal split was noted on the catheter body.Upon infusion, a leak was observed from the longitudinal split.Furthermore, a complete diagonal break was noted on the distal end.However, striations was noted throughout the catheter surface.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 08/2022), g3, h6 (device) h11: h6 (result, conclusion) h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event or Problem Description
It was reported that sometimes post port placement, the line allegedly found to have a leak.It was further reported that catheter allegedly had hole.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROVIAC CV CATHETER, SINGLE-LUMEN, 2.7F
Common Device Name
CHRONIC CATHETERS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15444388
Report Number3006260740-2022-03645
Device Sequence Number4750872
Product Code LJS
UDI-Device Identifier00801741036309
UDI-Public(01)00801741036309
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K830256
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0600040
Device Catalogue Number0600040
Device Lot NumberREEP4290
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/23/2022
Supplement Date Received by Manufacturer11/29/2022
Initial Report FDA Received Date09/19/2022
Supplement Report FDA Received Date11/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age17 DA
Patient SexFemale
Patient Weight805 KG
-
-