C.R. BARD, INC. (BASD) -3006260740 POWERPORT CLEARVUE ISP IMPLANTABLE PORT, CHRONOFLEX, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 1668362
Device Problems Device Appears to Trigger Rejection (1524); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Thrombosis/Thrombus (4440); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/24/2022
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiration date: 09/2023).Device pending return.
 
Event Description
It was reported that some time post port placement procedure, patient allegedly experienced thrombosis in jugular vein.However, at this point of time there was no evidence that the device caused thrombosis.There was no reported patient injury.
 
Manufacturer Narrative
H10: the initial mdr was inadvertently submitted with a g3 date of 08/23/2022.The correct g3 date is 08/24/2022.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.Furthermore, the clinical conditions reported in the complaint cannot be confirmed.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 09/2023), g3, h6 (method).H11: b3, h6 (result, conclusion).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that some time post port placement procedure, patient allegedly experienced thrombosis in jugular vein.However, at this point of time there was no evidence that the device caused thrombosis.There was no reported patient injury.
 
Manufacturer Narrative
H10: based on the review of the event information, the reportability of the event has been upgraded.Therefore, the file was reassessed for reportability and determined to be reportable as serious injury.H10: as the lot number for the device was provided, a review of the device history records was be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiration date: 09/2023), g3.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that sometime after port placement procedure, patient allegedly experienced thrombosis.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT CLEARVUE ISP IMPLANTABLE PORT, CHRONOFLEX, 8F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15445265
MDR Text Key303554286
Report Number3006260740-2022-03647
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741110740
UDI-Public(01)00801741110740
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063377
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/19/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1668362
Device Catalogue Number1668362
Device Lot NumberREGR4458
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/09/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-