• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. CLAV FRAC PLT,CENTRAL THIRD,LFT,SS; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. CLAV FRAC PLT,CENTRAL THIRD,LFT,SS; PLATE, FIXATION, BONE Back to Search Results
Model Number CLAV FRAC PLT,CENTRAL THIRD,LFT,SS
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 08/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that after an initial implantation of the plate on (b)(6) 2022 the plate broke post-op without external forces.The patient went to the hospital and received the diagnosis that the plate broke on (b)(6) 2022 and the revision surgery was then performed on (b)(6) 2022.
 
Additional Manufacturer Narrative
Complaint confirmed.Upon visual inspection, it was noted that the ar-2651cl plate was broken within the 5th screw slot, between the ce mark and the batch number.The width and thickness of the plate were measured and found to be within tolerance.The cause of this remains undetermined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAV FRAC PLT,CENTRAL THIRD,LFT,SS
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15445820
Report Number1220246-2022-05503
Device Sequence Number20003400
Product Code HRS
UDI-Device Identifier00888867027404
UDI-Public00888867027404
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K143139
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberCLAV FRAC PLT,CENTRAL THIRD,LFT,SS
Device Catalogue NumberAR-2651CL
Device Lot Number105642003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2022
Supplement Date Received by Manufacturer08/31/2022
Initial Report FDA Received Date09/19/2022
Supplement Report FDA Received Date12/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/28/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-