Model Number M0036122060
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the procedure, the coil (subject device) felt resistance within the microcatheter and was unable to be deployed.The physician replaced it with a new device and continued the procedure without clinical consequences to the patient.The subject device was returned for analysis and it was discovered that the main coil (subject device) was found to be broken/fractured during use.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.During visual inspection, the main coil was found to be stretched, tangled and broken/fractured.The delivery wire was found to be kinked/bent.A functional test for the reported event 'coil in catheter friction' was unable to be performed due to the condition of the device.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported event could not be confirmed; however, the analysis results are consistent with the reported event.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.The device was analyzed.The main coil was found to be stretched, tangled and broken/fractured.This complaint appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.Therefore, an assignable cause of procedural factors will be assigned to the ar code 'coil in catheter friction' and to the aa codes 'main coil stretched', 'main coil broken/fractured during use', and 'main coil tangled.' an assignable cause of 'handling damage' will be assigned to aa code 'coil delivery wire kinked/bent' as these defects appear to be associated with handling of the product or portion of the product during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XL 360 SOFT 2MM X 6CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15446434
Report Number3008881809-2022-00458
Device Sequence Number4750407
Product Code HCG
UDI-Device Identifier07613327128260
UDI-Public07613327128260
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/08/2024
Device Model NumberM0036122060
Device Catalogue NumberM0036122060
Device Lot Number23086407
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/05/2022
Initial Report FDA Received Date09/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-