| Model Number |
N/A |
| Medical Device Problem Codes |
Difficult to Insert (1316); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Implanted date: device was not implanted.Explanted date: device was not explanted.Initial reporter: occupation: cardio lab manager.The actual device was not returned for evaluation.The investigation is currently ongoing.A follow-up report will be submitted once the investigation is complete.A review of the device history record of the product code/lot# combination was conducted with no findings.
|
| |
|
Event or Problem Description
|
|
The user facility reported that the wire seemed to have a prominent weld that prevented the wire from passing through the needle.There was no patient injury, or blood loss.The procedure was successful, and the patient was in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow-up no.1 to update section d9 and h3, and to provide the completed investigation results.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.The complaint cannot be confirmed for guidewire and needle mobility issues because the sample was not returned for assessment.The exact root cause could not be determined.Review of device history record (dhr) showed there were no issues noted during manufacture of the product that could have contributed to this complaint condition.Currently no action is recommended since this risk evaluation is within the predetermined limits in the design failure mode & effects analysis (fmea).
|
| |
|
Search Alerts/Recalls
|