• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE SG; PERMANENT DEFIBRILLATOR ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE SG; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number 0283
Medical Device Problem Codes High impedance (1291); High Capture Threshold (3266)
Health Effect - Clinical Code Shock (2072)
Date of Event 01/19/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device remains implanted.As such, physical analysis has not been conducted in our laboratory.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that this right ventricular (rv) lead exhibited high, out of range shock impedance, rising since november 2021 from 125 ohms to 175 ohms.In 2019, device delivered shock therapy and shock impedance was at 65 ohms.Rv pacing impedance increased from 800 ohms ( at implant) to 1700 ohms.Rv pacing threshold where 0.5ms x 0.5ms (at implant) to 2.7ms x 1.0ms reported today.The physician will continue to monitor the patient over the home monitoring system.The rv lead remains implanted.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
Device remains implanted.As such, physical analysis has not been conducted in our laboratory.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that this right ventricular (rv) lead exhibited high, out of range shock impedance, rising since november 2021 from 125 ohms to 175 ohms.In 2019, device delivered shock therapy and shock impedance was at 65 ohms.Rv pacing impedance increased from 800 ohms (at implant) to 1700 ohms.Rv pacing threshold were 0.5ms x 0.5ms (at implant) to 2.7ms x 1.0ms reported today.The physician will continue to monitor the patient over the home monitoring system.A technical service (ts) consultant review device data to provide recommendations for additional testing.Including lead integrity testing with provocation maneuvers/isometrics, x-ray images, pocket manipulation and commanded 1.1j and 41j shocks.Additional information was received that lead integrity testing was completed.A commanded 1.1j shock and 41j shock where completed.Both revealed shock impedance of 120 ohms.Induction testing was completed, and shock impedance was evaluated at 105 ohms through induction testing.The physician will continue monitoring the patient closely.The rv lead remains implanted.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE SG
Common Device Name
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15447203
Report Number2124215-2022-36648
Device Sequence Number4799960
Product Code NVY
UDI-Device Identifier00802526431388
UDI-Public00802526431388
Combination Product (Y/N)Y
Initial Reporter CountryIT
PMA/510(K) Number
P910073/S077
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/11/2015
Device Model Number0283
Device Catalogue Number0283
Device Lot Number101432
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/13/2022
Supplement Date Received by Manufacturer10/21/2022
Initial Report FDA Received Date09/19/2022
Supplement Report FDA Received Date10/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/11/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-