EXACTECH, INC. LGC TIBIAL FIT TRAY CEM SZ 3F / 3T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number LOGIC CR TIB INSERT STD, SZ 3, 9MM
Medical Device Problem Codes Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Osteolysis (2377)
Date of Event 08/11/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, the patient had an initial right tka on (b)(6) 2019.The patient was revised on (b)(6) 2022 due to wear of the insert.The patient was last known to be in stable condition following the event.No other medical history/patient information provided.
 
Additional Manufacturer Narrative
Pending investigation.Concomitant medical products: 02-010-03-0330 - logic cr femoral cem, right, sz 3 (b)(4).02-012-45-3030 - lgc tibial fit tray cem sz 3f / 3t (b)(4).1400-ag - cemex gento low viscosity 40g kit ab3521.200-02-35 - three peg patella 35mm (b)(4).204-32-01 - fluted stem extension 11l x 12 mm (b)(4).204-70-00 - tibial stem ext.Screw (b)(4).Recall: z-0021-2022.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The reason for the revision reported cannot be confirmed from the information provided but may be the result of tibial loosening, prosthesis wear, and patellar loosening.Or due to inclusion of the polyethylene in the packaging recall.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGC TIBIAL FIT TRAY CEM SZ 3F / 3T
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key15447759
Report Number1038671-2022-01135
Device Sequence Number4622562
Product Code JWH
UDI-Device Identifier10885862003966
UDI-Public10885862003966
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K111400
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 01/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberLOGIC CR TIB INSERT STD, SZ 3, 9MM
Device Catalogue Number02-012-45-3030
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2022
Supplement Date Received by Manufacturer01/16/2025
Initial Report FDA Received Date09/19/2022
Supplement Report FDA Received Date01/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/01/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-