• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. TRULIANT CR CEM FEM CR CEM RIGHT SZ 3; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. TRULIANT CR CEM FEM CR CEM RIGHT SZ 3; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Model Number TRULIANT CR CEM FEM CR CEM RIGHT SZ 3
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Emotional Changes (1831); Failure of Implant (1924); Pain (1994); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Balance Problems (4401); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); Joint effusion (4849)
Date of Event 08/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending investigation.Concomitant medical products: (b)(4), 200-07-32 - advanced patella 32mm 3 peg implant.(b)(4), 02-022-51-3010 - truliant tib imp crc insert sz 3, 10mm.(b)(4), 02-022-45-3020 - truliant tib fit tray cem sz 3f / 2t.
 
Event or Problem Description
As reported, the patient had an initial right tka on (b)(6) 2018.The patient complained of pain and was revised due to a loose femur.The patient was last known to be in stable condition following the event.No other medical history/patient information provided.
 
Additional Manufacturer Narrative
H3: the revision reported was likely the result of aseptic loosening between the femoral component and the bone.However, this cannot be determined as the devices were not available for evaluation and limited clinical information was provided.H7: z-0023-2022.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Mdr section codes updated/corrected: e, g.The revision reported was likely the result of aseptic loosening between the femoral component and the bone.However, this cannot be determined as the devices were not available for evaluation and limited clinical information was provided if any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT CR CEM FEM CR CEM RIGHT SZ 3
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key15447826
Report Number1038671-2022-01136
Device Sequence Number17325896
Product Code JWH
UDI-Device Identifier10885862314673
UDI-Public10885862314673
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K170240
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberTRULIANT CR CEM FEM CR CEM RIGHT SZ 3
Device Catalogue Number02-020-13-0330
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2022
Supplement Date Received by Manufacturer03/15/2023
02/20/2025
03/28/2025
Initial Report FDA Received Date09/19/2022
Supplement Report FDA Received Date03/01/2023
02/25/2025
03/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/31/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexFemale
-
-