MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ASCENDA; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ASCENDA; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8781
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Pain (1994)
Date of Event 08/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from healthcare provider (hcp) via manufacturer representative (rep) regarding a patient receiving fentanyl (2500 mcg ml/320 mcg day) via an implantable pump.It was reported that the catheter was not working.The patient healthcare provider (hcp) tried to aspirate through the side port upon the patient complaining he was no longer getting adequate pain relief.She was not able to obtain fluid through the side port.The patient hcp scheduled surgery for (b)(6) 2022 and tried to aspirate from the pump side port in surgery and was also not able to aspirate so he explanted the old catheter and implanted a new 8781 ascenda catheter.The old catheter was x planted in a new catheter was implanted.There was no environmental factor/patient factor.The patient¿s medical history and weight was asked but unknown.The patient status was alive-no injury.The issue was resolved.
 
Event or Problem Description
Information was received from healthcare provider (hcp) via manufacturer representative (rep) regarding a patient receiving fentanyl (2500 mcg ml/320 mcg day) via an implantable pump.It was reported that the catheter was not working.The patient healthcare provider (hcp) tried to aspirate through the side port upon the patient complaining he was no longer getting adequate pain relief.She was not able to obtain fluid through the side port.The patient hcp scheduled surgery for (b)(6) 2022 and tried to aspirate from the pump side port in surgery and was also not able to aspirate so he explanted the old catheter and implanted a new 8781 ascenda catheter.The old catheter was x planted in a new catheter was implanted.There was no environmental factor/patient factor.The patient¿s medical history and weight was asked but unknown.The patient status was alive-no injury.The issue was resolved.
 
Additional Manufacturer Narrative
Continuation of d10: product id 8637-40 serial# (b)(6) implanted: (b)(6) 2021: product type pump it was accidentally reported on the catheter before.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASCENDA
Common Device Name
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key15449106
Report Number3004209178-2022-12045
Device Sequence Number18020263
Product Code LKK
UDI-Device Identifier00763000051129
UDI-Public00763000051129
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/16/2022
Device Model Number8781
Device Catalogue Number8781
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2022
Supplement Date Received by Manufacturer09/13/2022
Initial Report FDA Received Date09/19/2022
Supplement Report FDA Received Date09/19/2022
Date Device Manufactured11/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H10...."
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexMale
-
-