| Model Number |
1365-32-710 |
| Medical Device Problem Code |
Noise, Audible (3273)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Surgeon reported to bfarm: "recurrence of inlay dislocation after tep implantation on (b)(6) 2021.First dislocation of the inlay on (b)(6) 2022, on (b)(6) 2022 change of the inlay, on (b)(6) 2022 another inlay dislocation occurred during deep bending.Intraoperatively, no inlay could be inserted in a stabile manner in the implanted cup.".
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary :the device associated with this report was received for examination.Visual examination of the returned femoral head denoted a remarkable worn appearance at the part.It was possible to identify a disassociation event among poly liner and metal cup since it was found that occurred a transfer of the titanium cup material onto the surface of the ceramic head.It is not unreasonable that noise would be present due to the ceramic head articulating against the metal cup.It is subsequent the liner disassociation and does not signify any product error.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : quantity manufactured: (b)(4).2) date of manufacture: 02/2019.3) any anomalies or deviations identified in dhr: no non-conformances nor deviations.4) expiry date: 01/2024.5) ifu reference: (b)(4).Device history review : a manufacturing record evaluation was performed for the finished device (136532710/9051220) product and lot numbers, and no non-conformances were identified.H10 additional narrative: added: h4 corrected: h3.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary the device associated with this report was received for examination.Visual examination of the returned femoral head denoted a remarkable worn appearance at the part.It was possible to identify a disassociation event among poly liner and metal cup since it was found that occurred a transfer of the titanium cup material onto the surface of the ceramic head.It is not unreasonable that noise would be present due to the ceramic head articulating against the metal cup.It is subsequent the liner and cup disassociation and does not signify any product error.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot 1) quantity manufactured: (b)(4).2) date of manufacture: 02/14/2019.3) any anomalies or deviations identified in dhr: no non-conformances nor deviations.4) expiry date: 01/2024.5) ifu reference: (b)(4).Device history review a manufacturing record evaluation was performed for the finished device (136532710/9051220) product and lot numbers, and no non-conformances were identified.
|
| |
|