MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 802018
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The field service representative (fsr) verified the reported complaint.The light emitting diode (led) on the module did not come on.The fsr replaced the assembly flow module.The unit operated to the manufacturer's specifications.
 
Event or Problem Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the central control monitor (ccm) displayed a 'service module art flow' message.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Additional Manufacturer Narrative
Updated block: h6 the reported complaint was confirmed.Per supplier evaluation, the five volt (v) and ground were shorted.This was confirmed with a digital voltmeter.After removing components one by one it was found that the integrated circuit at the u11 position had failed causing the short.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
Updated blocks: d9 and h6.During laboratory analysis, the product surveillance technician (pst) installed the flow module into lab use only (luo) testing equipment.Upon powering on the unit, the light emitting diode (led) did not power on and the central control monitor (ccm) displayed a 'flow1: service module' message.The pst replaced the printed circuit board assembly (pcba) from the module with a luo pcba.The module then functioned as intended determining that the pcba of the module was defective.Upon microscopic evaluation of the pcba, he had found that the electrical component u1 was damaged and failed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
Per data log review: on (b)(6) 2022 the flow meter reported several errors.5.0 volt (v) supply range status event (8) 7.5v supply range status event (8) motor voltage (vmot) power state change (disabled) vmot over-current alarm (710) pod initialization failure - erasable electronically programmable read only memory (eeprom) checksum invalid this would have caused the message 'art flow: service module' as reported.Subsequent power ups caused the pod initialization failure - eeprom checksum invalid each time.The log confirms the complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Common Device Name
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key15451127
Report Number1828100-2022-00334
Device Sequence Number5835904
Product Code DTQ
UDI-Device Identifier00886799000687
UDI-Public(01)00886799000687(11)220502
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K172220
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 12/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number802018
Device Catalogue Number802018
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2022
Supplement Date Received by Manufacturer09/19/2022
10/11/2022
12/21/2022
Initial Report FDA Received Date09/19/2022
Supplement Report FDA Received Date10/06/2022
11/01/2022
12/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/02/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-