• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX TORIC TRIFOCAL IOL; LENS, MULTIFOCAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX TORIC TRIFOCAL IOL; LENS, MULTIFOCAL Back to Search Results
Model Number TFNT30
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Eye Pain (4467)
Date of Event 06/04/2021
Type of Reportable Event Serious Injury
Event or Problem Description
A non healthcare professional reported following an intraocular lens implant procedure, the patient experienced poor best corrected visual acuity and intolerable asthenopia.The surgeon indicated over correction of 1 diopter of cylinder, therefore the lens was exchanged in a secondary procedure.Additional information has been requested.
 
Additional Manufacturer Narrative
The lens was returned for evaluation.Solution is dried on the lens.The lens is cut into two pieces.One piece has marks on the optic in the shape of a surgical instrument used to grasp the lens.All product and batch history records are quality reviewed prior to product release.Qualified associated products were indicated.The product investigation could not identify a root cause for the reported complaint.Information was provided that in the surgeon's opinion the cause of the even was post op indication of over correction of 1 diopter of cylinder.There are three other complaints in this lot.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX TORIC TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15451158
Report Number1119421-2022-01997
Device Sequence Number5581977
Product Code MFK
UDI-Device Identifier00380652415484
UDI-Public00380652415484
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model NumberTFNT30
Device Catalogue NumberTFNT30U180
Device Lot Number12700483
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/25/2022
Initial Report FDA Received Date09/19/2022
Date Device Manufactured04/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DUOVISC OVD; MONARCH III IOL CARTRIDGE D; MONARCH III IOL, INJECTOR
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexMale
-
-