• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 PPS LTD ACET SHELL 54F; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 PPS LTD ACET SHELL 54F; PROSTHESIS, HIP Back to Search Results
Catalog Number 010000664
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the lock on the impactor cross threaded into the cup and the drive broke while trying to disengage.There were no health consequences or impact to the patient.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).Concomitant medial products: cat #:010002736 - g7 ball hex drvr for insr hndl cat #: 110003451 - g7 str modular shell inserter lot #: 503420.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2022-02164, 0001825034-2022-02165.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No additional event information to report at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected.Updated: b4; b5; d4; g3; h2; h3; h4; h6.Visual examination of the returned product identified the devices were stuck together so no functional analysis was completed.There was an indentation on the shaft of the inserter and the tip has a roll on the threads.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 PPS LTD ACET SHELL 54F
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15451173
Report Number0001825034-2022-02166
Device Sequence Number5582479
Product Code PBI
UDI-Device Identifier00880304524231
UDI-Public(01)00880304524231(17)320630(10)7286498
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number010000664
Device Lot Number7286498
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2022
Supplement Date Received by Manufacturer10/03/2022
Initial Report FDA Received Date09/19/2022
Supplement Report FDA Received Date10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Patient Age78 YR
Patient SexFemale
-
-