Model Number 7209739
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/30/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during an unspecified procedure, when the drilling in the femoral tunnel began, the 10mm flexible drill device broke.It is unknown whether if there was a back-up device available or if there was a delay.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during an acl reconstruction procedure, when the drilling in the femoral tunnel began, the 10mm flexible drill device broke.All broken pieces were removed using a hemostat.The procedure was successfully completed with non-significant delay using a flexible 9.5mm auger.Patient is reported to be recovering normally and no further complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.An analysis of the customer provided image found that the tip of the drill was missing from the shaft.The root cause of the reported event could not be determined.Factors that could have contributed to the failure include excessive force on the device, excessive torque on the device tip, attempted correction of a damaged device, or an inadvertent impact event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL FLEXIBLE ENDOSCOPIC CANN 10MM
Common Device Name
BIT, DRILL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15451852
Report Number1219602-2022-01388
Device Sequence Number6020782
Product Code HTW
UDI-Device Identifier03596010541970
UDI-Public03596010541970
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7209739
Device Catalogue Number7209739
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2022
Supplement Date Received by Manufacturer10/11/2022
10/27/2023
Initial Report FDA Received Date09/19/2022
Supplement Report FDA Received Date10/12/2022
11/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-