• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GN HEARING A/S GN RESOUND; HEARING AID, AIR CONDUCTION WITH WIRELESS TECHNOLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GN HEARING A/S GN RESOUND; HEARING AID, AIR CONDUCTION WITH WIRELESS TECHNOLOGY Back to Search Results
Model Number 19706708
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
Sf: (b)(4).It was reported- aids put in dry jar and melted.Customer verified the jar was not resound brand.They believe the heat element in the jar caused the melting and therefore was not related to the resound aids, but no specific investigation was done' the current information indicates need for immediate reporting.Based on the available information, the event/ incident did lead to, or might led to, death or a serious injury if it were to recur.This is an initial report.
 
Additional Manufacturer Narrative
Sf: (b)(4).A dhr review have been completed.No deviation or changes were found during manufacturing of the device.It has been reported- not enough information how the ha got melted in the dry jar.Aid physically melted and broken into pieces.Technical investigation: most drying jars can dry aids in surroundings up to 50-60 degrees.If the hearing aids are turned off, it is fine to use dry jars.A dry box can be used for all types of gn hearing aids.It is suspected that the aids were left in the dry box subjected to surroundings with excessive heat.This is an initial report.
 
Event or Problem Description
(b)(6).It was reported- aids put in dry jar and melted.Customer verified the jar was not resound brand.They believe the heat element in the jar caused the melting and therefore was not related to the resound aids, but no specific investigation was done'.The current information indicates need for reporting.Based on the available information, the event/ incident did lead to, or might led to, death or a serious injury if it were to recur.This is a final report.
 
Additional Manufacturer Narrative
(b)(6).A dhr review have been completed.No deviation or changes were found during manufacturing of the device.It has been reported- not enough information how the ha got melted in the dry jar.Aid physically melted and broken into pieces.Technical investigation: most drying jars can dry aids in surroundings up to 50-60 degrees.If the hearing aids are turned off, it is fine to use dry jars.A dry box can be used for all types of gn hearing aids.It is suspected that the aids were left in the dry box subjected to surroundings with excessive heat.Clinical evaluation of the event: case information states that the devices melted while placed in a drying jar.The jar is not a gn hearing product, and the patient reports that the heating element in the drying jar may have caused the melting.No patient harm was reported, as the devices were in the drying jar and not in the patient's ears.Device user guide has a warning how to dry the devices.Risk analysis already included, the residual risk after the risk control measure(s) is acceptable.This is a final report.No further information is expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GN RESOUND
Common Device Name
HEARING AID, AIR CONDUCTION WITH WIRELESS TECHNOLOGY
Manufacturer (Section D)
GN HEARING A/S
lautrupbjerg 7
ballerup, 2750
DA  2750
Manufacturer Contact
karen frisk
lautrupbjerg 7
ballerup, 2750
DA   2750
MDR Report Key15452954
Report Number3005650109-2022-00035
Device Sequence Number20551418
Product Code OSM
UDI-Device Identifier05704819289933
UDI-Public05704819289933
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number19706708
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2022
Supplement Date Received by Manufacturer08/22/2022
Initial Report FDA Received Date09/20/2022
Supplement Report FDA Received Date10/18/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/08/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-