Model Number SV25T0
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Corneal Edema (1791); Blurred Vision (2137); Visual Disturbances (2140)
Date of Event 01/28/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A health care professional reported that following a cataract extraction with intraocular lens (iol) implant procedure, the patient was unhappy with vision and diplopia.Post surgery, the patient experienced corneal edema and vision was getting worse over time.The iol was explanted in a secondary procedure.Additional information has been requested and received stated that the patient issue has been resolved and the lens was exchanged for a non company lens.No further information available.There are two medical device reports associated with this patient.This report is associated with the left eye.
 
Additional Manufacturer Narrative
The lens was returned in a specimen cut.Solution was dried on the lens.The optic was torn/split/cracked and cut into pieces, typical of insertion and removal.Power and resolution testing could not be conducted due to the extensive optic damage.Power and resolution testing could not be conducted due to the extensive optic damage.The power and resolution of each lens is 100% evaluated during the manufacturing process to determine acceptability per model and diopter.Information was provide that the multifocal company lens 20.0 diopter, +2.50 add power lens model was exchanged for a non-company 19.5 diopter monofocal lens.Due to the exchange of a multifocal lens to a monofocal lens may suggest that the replacement lens model was an improved choice for the patient's vision needs.No additional information has been provided at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ RESTOR SINGLEPIECE IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15453383
Report Number1119421-2022-01999
Device Sequence Number20585790
Product Code MFK
UDI-Device Identifier00380652307741
UDI-Public00380652307741
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Model NumberSV25T0
Device Catalogue NumberSV25T0U200
Device Lot Number12436336
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/25/2022
Supplement Date Received by Manufacturer10/11/2022
Initial Report FDA Received Date09/20/2022
Supplement Report FDA Received Date11/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/25/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age81 YR
Patient SexFemale
Patient Weight59 KG
-
-