MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. FLOWMATE PROCEDURAL KIT; CTO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. FLOWMATE PROCEDURAL KIT; CTO Back to Search Results
Model Number INJKIT
Medical Device Problem Codes Device Markings/Labelling Problem (2911); Inaccurate Information (4051)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history records could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.
 
Event or Problem Description
It was reported that prior to a recanalization procedure, the expiration date in kit was allegedly not the same from date on the box.There was no patient contact.
 
Event or Problem Description
It was reported that prior to a recanalization procedure, the expiration date in kit was allegedly not the same from date on the box.There was no patient contact.
 
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a photo were provided for review.The investigation of the reported event is currently underway.H10: d4(expiry date: 07/2022), g3.H11: h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.One electronic photo was provided and reviewed.The photo shows the flowmate procedural kit boxes and the components of the flowmate procedural kit.Per the drawing for the flowmate procedural kit, this is a kitted product and the original manufacturing labels do not get over labeled; the facility that packages the components only places a label on the master carton (outside box).The user should reference the label on the master carton for expiration date purposes.Therefore, the investigation is unconfirmed for the labeling issue.A definitive root cause for the alleged labeling issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 07/2022), g3, h6 (method).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that prior to a recanalization procedure, expiration date in the kit was allegedly not matching with the date on the box.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOWMATE PROCEDURAL KIT
Common Device Name
CTO
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15453457
Report Number2020394-2022-00747
Device Sequence Number5948709
Product Code DXT
UDI-Device Identifier00801741067693
UDI-Public(01)00801741067693
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K903390
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/29/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberINJKIT
Device Catalogue NumberINJKIT
Device Lot NumberREFT2431
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/01/2022
Supplement Date Received by Manufacturer09/22/2022
12/23/2022
Initial Report FDA Received Date09/20/2022
Supplement Report FDA Received Date10/19/2022
01/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-