Model Number EX061703CS
Medical Device Problem Codes Difficult to Remove (1528); Detachment of Device or Device Component (2907); Device-Device Incompatibility (2919); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation and images were provided for review.The investigation of the reported event is currently underway.Expiry date: 02/2024.
 
Event or Problem Description
It was reported that during a stent placement procedure in the right superficial femoral artery via contralateral approach, the stent allegedly foreshortened.The procedure was completed using another device.There was no reported patient injury.
 
Event or Problem Description
It was reported that during a stent placement procedure in the right superficial femoral artery via contralateral approach, the stent allegedly foreshortened.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the stent delivery system was returned for evaluation and was found in used condition without stent, which was completely deployed.The delivery sheath of the system was found stuck inside a device compatible 6f introducer sheath and loaded over a device compatible 0.035 inch guidewire.A joint inside the delivery system was found detached.Based on sample condition no indication for stent foreshortening could be found; x-ray images were provided, but with poor, dotty resolution so that a strut evaluation or length measurement was not possible, which leads to inconclusive result for stent foreshortening.The investigation will be closed as confirmed for detachment, difficult removal, and difficulties to remove the guidewire and introducer sheath.Based on the information available the investigation is inconclusive for stent foreshortening and confirmed for difficulty removal, detachment, and device incompatibility.A definite root cause for the reported event could not be determined based upon the available information.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.The instructions for use closely describes holding and handling of the system during deployment, in particular the instructions for use state: 'remove slack from the stent system held outside the patient.', 'confirm that the introducer sheath is secure and will not move during deployment.', and 'to ensure the most accurate placement, firmly hold the black system stability sheath throughout deployment.Note: do not hold the silver stent delivery sheath at any time during deployment.Do not constrict the stent delivery sheath during stent deployment.' in regards to access and accessories the instructions for use state: 'gain femoral access at the appropriate site using a 6f (2.0 mm) or larger introducer sheath (.) insert a 0.035¿ diameter guidewire of appropriate length'.The instructions for use further state: predilation of the lesion should be performed using standard techniques.', and 'if resistance is met during stent system introduction, the stent system should be removed and another stent system should be used.', and 'if resistance is met while retracting the delivery system over a guidewire, remove the delivery system and guidewire together.' h10: d4 (expiry date: 02/2024), g3, h6 (device, method).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTENT VASCULAR STENT
Common Device Name
VASCULAR STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15453790
Report Number9681442-2022-00286
Device Sequence Number17868157
Product Code NIP
UDI-Device Identifier04049519000961
UDI-Public(01)04049519000961
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P070014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEX061703CS
Device Catalogue NumberEX061703CS
Device Lot NumberANGP1396
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/31/2022
Supplement Date Received by Manufacturer10/10/2022
Initial Report FDA Received Date09/20/2022
Supplement Report FDA Received Date10/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age77 YR
Patient SexFemale
Patient Weight68 KG
-
-