• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH TORQ LIMITING HANDLE - 4.5MM QK COUPLING; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH TORQ LIMITING HANDLE - 4.5MM QK COUPLING; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 03.835.043
Medical Device Problem Code Calibration Problem (2890)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported on unknown date that , during evaluation at service and repair it was found out that the torque limiting handle quick coupling part failed torque test low.There was no patient involvement.This report is for (1) torq limiting handle: 4.5mm qk coupling.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.The manufacture date of this item is 13-apr-2017.The service history review is unconfirmed.The customer reported the device p/n: 03.835.043 needs repair.The repair technician reported the device failed high.The repair technician determined the device needed to be disposed as it was found to be out of specification.The evaluation was confirmed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORQ LIMITING HANDLE - 4.5MM QK COUPLING
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15453949
Report Number8030965-2022-07088
Device Sequence Number12481128
Product Code LXH
UDI-Device Identifier10705034817569
UDI-Public(01)10705034817569
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.835.043
Device Catalogue Number03.835.043
Device Lot NumberH253256-42
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/29/2022
Supplement Date Received by Manufacturer09/26/2022
Initial Report FDA Received Date09/20/2022
Supplement Report FDA Received Date10/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/13/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-