LUTONIX, INC LUTONIX 018 DRUG COATED DILATATION CATHETER; DRUG COATED PTA BALLOON CATHETER Back to Search Results
Model Number 9014
Medical Device Problem Codes Inflation Problem (1310); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 03/2025).
 
Event or Problem Description
It was reported that during an angioplasty procedure, the drug-coated balloon allegedly had inflation problem.It was further reported that the device twisted.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one lutonix 018 drug coated balloon catheter was received for evaluation.During visual evaluation, the entire length of the catheter was examined for damage revealing no anomalies.During functional evaluation, an in-house presto inflation device was used to inflate the balloon to 7 atm.The balloon was inflated and maintained pressure, then deflated without issue.No other functional testing was performed.One radiographic image was provided and reviewed.The image shows an angioplasty balloon inflated within the left superficial femoral artery.It appears almost completely inflated except for a small portion in which the balloon appears twisted.Although the balloon was inflated and deflated with no twisting during functional evaluation of the returned sample, the investigation is confirmed for the reported inflation problem and balloon twist, as twisting was noted on the balloon from the provided radiographic image review.A definitive root cause for the reported inflation problem and balloon twist could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the distal superficial femoral artery via left femoral artery, the drug-coated balloon allegedly had inflation problem.It was further reported that the balloon allegedly got twisted.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUTONIX 018 DRUG COATED DILATATION CATHETER
Common Device Name
DRUG COATED PTA BALLOON CATHETER
Manufacturer (Section D)
LUTONIX, INC
9409 science center dr
new hope 55428
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15454310
Report Number3006513822-2022-00032
Device Sequence Number6902891
Product Code ONU
UDI-Device Identifier00801741184710
UDI-Public(01)00801741184710
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P130024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9014
Device Catalogue NumberLX1813053005F
Device Lot NumberGFGN1384
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/29/2022
Supplement Date Received by Manufacturer11/02/2022
Initial Report FDA Received Date09/20/2022
Supplement Report FDA Received Date11/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age63 YR
Patient SexMale
-
-