ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY EXTENDED VISION IOL; INTRAOCULAR LENS Back to Search Results
Model Number DFT015
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There has been one other complaint reported in the lot.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A health care professional reported that during the cataract surgery with intraocular lens (iol) implant procedure, the lens was implanted, but inspection under the microscope revealed a scratch.A lens change was carried out without negative effects on the patient.Additional information has been requested.
 
Additional Manufacturer Narrative
The lens was returned positioned incorrectly posterior side up in a lens case.Solution was dried on the lens.One haptic was broken in the distal area (not returned).The optic was pressed against a post and leaning down into the well area of the lens case.The optic has a scratch near the center of the lens and along the optic edge.Qualified associated products were indicated.Upon return, the lens was observed to be placed into the lens case incorrectly resulting in damage.The reported scratch was observed.Other lens damage was observed that may have occurred due to how the lens was positioned in the lens case for return.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ VIVITY EXTENDED VISION IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15454408
Report Number1119421-2022-02002
Device Sequence Number7011295
Product Code POE
UDI-Device Identifier00380652421263
UDI-Public00380652421263
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/03/2026
Device Model NumberDFT015
Device Lot Number15244531
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/23/2022
Supplement Date Received by Manufacturer10/17/2022
Initial Report FDA Received Date09/20/2022
Supplement Report FDA Received Date11/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH CARTRIDGE D; MONARCH INJECTOR; UNSPECIFIED VISCOSURGICAL DEVICE
Patient SexMale
-
-