H10: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing was performed.Visual inspection found no damage, kinks or other defects were found.Functional testing performed based on the no disposable alarm investigation and root cause conduct trough capa-000814 the mohawk exposure could be a factor during the use of cassette to activate the alarm.However, the failure mode could not be duplicated by functional testing and no mohawk exposure detected, complaint is not confirmed.Device passed all functional testing.No problem found.No actions taken, however as part of in process controls add based on capa opened on 09/sep/2021 for no disposable alarm issue, root cause the following containment action was implemented: work instructions updated from revision 104 to revision 105 to add the mohawk exposure criteria.Complete on 08/sep/2022.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|