• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH-CT SOMATOM DEFINITION AS; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH-CT SOMATOM DEFINITION AS; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number 08098027
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Power Problem (3010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a hardware error.A siemens service engineer evaluated the system onsite.The specialist found a defective vertical drive unit and rotation encoder.Both have been replaced and the system is back to normal operation.The material consumption of used replacements in relation to the installed base is monitored by the capa process.A possible error accumulation or even a systematic error, which would lead to a corrective action of the installed base, could not be determined by the investigation.No general design issue was identified.Therefore, this report has been submitted with an abundance of caution.
 
Event or Problem Description
It was reported to siemens that a malfunction occurred while operating the somatom definition as ct system.On (b)(6), 2022, a 48 year old male patient came in as an emergency patient with a cerebral hemorrhage and in a coma.The patient handling system (phs - patient table) was lifting with the patient on it when the gantry powered off.It was reported that this caused a delay in diagnosis of approximately 10 minutes.Currently the patient is still in the intensive care unit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOMATOM DEFINITION AS
Common Device Name
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH-CT
siemensstr. 1 -or-
rittigfeld 1
forchheim, germany 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH-CT
siemensstr. 1 -or-
rittigfeld 1
forchheim, germany 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
malvern, PA 19355
4843231631
MDR Report Key15454736
Report Number3004977335-2022-42061
Device Sequence Number15160337
Product Code JAK
UDI-Device Identifier04056869003665
UDI-Public04056869003665
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K190578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Remedial Action Repair,Inspection
Type of Report Initial
Report Date (Section B) 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number08098027
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/25/2022
Initial Report FDA Received Date09/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age48 YR
Patient SexMale
-
-