| Model Number |
N/A |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Loss of Range of Motion (2032)
|
| Date of Event |
08/17/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical products: 42500006001 - femur cemented ps - 64121832, 42512600510 - articular surface fixed bearing cps - 64163759, 00579104100 - headed screw 48 mm length - 64243320, 00579104100 - headed screw 48 mm length - 64243320, 42509902525 - 2.5 mm female hex screw 25 mm length - 64346440, unknown simplex hv cement.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2022-00244.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a left knee revision surgery approximately 3.5 years post implantation due to loosening.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; g6; h1; h2; h3; h6; h10 no product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: follow up: pain 5 out of 10, wants revision due to persistent pain; reports moderate stiffness in am.Rom 0-120 degrees, normal stability.Revision: patella well fixed and remained, tibial component removed easily at the implant cement interface, femoral component removed with minimal bone loss.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|