Catalog Number 5000-01-02 |
Device Problem
Insufficient Cooling (1130)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/07/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
|
Event Description
|
It was reported that the arctic sun device was determined to be not cooling during field service.
|
|
Manufacturer Narrative
|
The reported issue was unconfirmed.The root cause of the reported issue could not be determined as the reported issue could not be reproduced.The device was evaluated upon receipt.The reported event is unconfirmed, dhr review is not required.The reported event is unconfirmed, labeling/packaging review is not required.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
|
Event Description
|
It was reported that the arctic sun device was determined to be not cooling during field service.Per follow up information received on 09sep2022, there was no patient involved.Bd technician ran functionality test on device onsite.Device passed, unconfirming not cooling.Customer sent the device back to service.
|
|