| Model Number |
901068-02 |
| Medical Device Problem Codes |
Flaked (1246); Peeled/Delaminated (1454); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
08/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complaint investigation underway and will be attached to this report.
|
| |
|
Event or Problem Description
|
|
The physician introduced a shepherd guide wire and jade balloon to angioplasty a segment of the anterior tibial artery (at).After deflation, the physician attempted to walk the balloon distally to treat the next diseased segment but resistance was experienced and the balloon would not cross.The physician attempted to push and pull the guide wire, however, resistance was felt when the guide wire was pulled back.Under fluoroscopy, the physician removed the guide wire and balloon together.The polymer coating of the guide wire was observed to be stripped and bunched, which prevented the balloon to be pushed distally.The physician used a non-csi guide wire to deliver the balloon catheter distally and finish the angioplasty.
|
| |
|
Event or Problem Description
|
|
The physician introduced a shepherd guide wire and jade balloon to angioplasty a segment of the anterior tibial artery (at).After deflation, the physician attempted to walk the balloon distally to treat the next diseased segment but resistance was experienced and the balloon would not cross.The physician attempted to push and pull the guide wire, however, resistance was felt when the guide wire was pulled back.Under fluoroscopy, the physician removed the guide wire and balloon together.The polymer coating of the guide wire was observed to be stripped and bunched, which prevented the balloon to be pushed distally.The physician used a non-csi guide wire to deliver the balloon catheter distally and finish the angioplasty.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint investigation underway and will be attached to this report.
|
| |
|
Search Alerts/Recalls
|