Model Number LTF-S300-10-3D |
Device Problems
No Display/Image (1183); Image Display Error/Artifact (1304); Electrical Shorting (2926)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
The distributor reported the contact on the endoeye flex 3d deflectable videoscope was poor.The image was not stable.No patient injury or procedure impact reported.During the evaluation of the device, it was noted an e216 (scope communication error) occurred due to a short-circuit of the image sensor cable.This report is to capture the reportable malfunction of the e216 (scope communication error) occurred due to a short-circuit of the image sensor cable noted at estimation.
|
|
Manufacturer Narrative
|
No additional information was available.Initial reporter name and address: (phone/email): additional information was requested.A supplemental report will be submitted should additional information be made available.The device was evaluated, and the reported issue was not confirmed.The bending section cover had scratches caused by external factors.The bending section cover adhesive was missing.The video connector was cracked due to external factors.The investigation was ongoing.A supplemental report will be submitted upon completion of the investigation.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified; however, the event most likely occurred due to a malfunction of the image sensor unit, a board inside the video connector, or a malfunction on the system side.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|