• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE; FLEX DEFLECTABLE VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE; FLEX DEFLECTABLE VIDEOSCOPE Back to Search Results
Model Number LTF-S300-10-3D
Device Problems No Display/Image (1183); Image Display Error/Artifact (1304); Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The distributor reported the contact on the endoeye flex 3d deflectable videoscope was poor.The image was not stable.No patient injury or procedure impact reported.During the evaluation of the device, it was noted an e216 (scope communication error) occurred due to a short-circuit of the image sensor cable.This report is to capture the reportable malfunction of the e216 (scope communication error) occurred due to a short-circuit of the image sensor cable noted at estimation.
 
Manufacturer Narrative
No additional information was available.Initial reporter name and address: (phone/email): additional information was requested.A supplemental report will be submitted should additional information be made available.The device was evaluated, and the reported issue was not confirmed.The bending section cover had scratches caused by external factors.The bending section cover adhesive was missing.The video connector was cracked due to external factors.The investigation was ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified; however, the event most likely occurred due to a malfunction of the image sensor unit, a board inside the video connector, or a malfunction on the system side.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15455425
MDR Text Key301328075
Report Number9610595-2022-01984
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-S300-10-3D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2022
Was the Report Sent to FDA? No
Date Manufacturer Received10/06/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/16/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-