• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. JUGGERSTITCH CURVED IMPLANT; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE/SPORTS MEDICINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. JUGGERSTITCH CURVED IMPLANT; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE/SPORTS MEDICINE Back to Search Results
Model Number 110024773
Device Problems Fracture (1260); Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/23/2022
Event Type  malfunction  
Event Description
It was reported that the surgeon had difficulty inserting the needle into the patient's meniscus.Once the surgeon was able to do this, the instrument failed to deploy the anchor and needle was detached from the device.No patient injury has been reported as a result of the malfunction.
 
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; g6; h1; h2; h3; h6 visual examination of the returned product identified the product has been cycled twice.The suture lines are free from the device and there is one suture line that has nothing attached and there is one that has the fractured needle tip and anchor.It is confirmed the needle tip has fractured and all the pieces were returned.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JUGGERSTITCH CURVED IMPLANT
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE/SPORTS MEDICINE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15455911
MDR Text Key302244174
Report Number0001825034-2022-02163
Device Sequence Number1
Product Code MBI
UDI-Device Identifier00880304690073
UDI-Public(01)00880304690073(17)270524(10)292870
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191459
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number110024773
Device Catalogue Number110024773
Device Lot Number292870
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age18 YR
Patient SexFemale
-
-