| Model Number |
50-7000B |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
Pneumothorax (2012)
|
| Date of Event |
09/02/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4) initial emdr submission.A follow up emdr will be submitted if additional information becomes available.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported by the customer that the piece from the product was not patent.Verbatim: an incident that happened around 5 pm, after surgery regarding a patient.The fellow came to the surgery front desk and informed of an incident.Patient had a tension pneumo following his video assisted thoracotomy with a pleurax placement.A white connector piece from the kit was not patent.He was trouble shooting the system at the patient's bedside of why the catheter was not working properly.When he disassemble the setup, he noticed the piece from the kit was not patent.This piece was immediately removed and changed and then the catheter was working properly.The fellow brought the connector back to the operating room for the surgical services to follow up.I called sicu to check on the patient he was stable and doing well.The lot number of the inserted catheter is 0001471891 we had 2 in the surgical unit that were pulled, and materials management was notified of the lot number to pull and retain all units to swap out for different kits.Fda safety report id# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Follow-up emdr for device evaluation: no photos or samples were received by our quality team for evaluation therefore the failure mode could not be verified.A review of the internal manufacturing device records and raw material history files for the reported lot number 0001471891 was performed and no recorded quality problems or rejections related to this incident were found.Product undergoes inspections during manufacturing, no issues related to the reported incident were identified, all procedural and functional requirements for product release have been met.Based on the available information we are not able to identify a root cause at this time.We appreciate you taking the time to bring this observation to our attention.Complaints received for this device and defect will continue to be monitored by our quality team for signs of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer that the piece from the product was not patent.Verbatim: an incident that happened around 5 pm, after surgery regarding a patient.The fellow came to the surgery front desk and informed of an incident.Patient had a tension pneumo following his video assisted thoracotomy with a pleurax placement.A white connector piece from the kit was not patent.He was trouble shooting the system at the patient's bedside of why the catheter was not working properly.When he disassemble the setup, he noticed the piece from the kit was not patent.This piece was immediately removed and changed and then the catheter was working properly.The fellow brought the connector back to the operating room for the surgical services to follow up.I called sicu to check on the patient he was stable and doing well.The lot number of the inserted catheter is 0001471891 we had 2 in the surgical unit that were pulled, and materials management was notified of the lot number to pull and retain all units to swap out for different kits.Fda safety report id# (b)(4).
|
| |
|