MICRO THERAPEUTICS, INC. DBA EV3 CONCERTO PGLA HLX; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
|
Back to Search Results |
|
| Model Number |
PV-3-8-3D |
| Medical Device Problem Codes |
Separation Failure (2547); Material Deformation (2976); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter.Treating physician's information was not reported.A separate report will be submitted regarding the other coil that failed to detach during the procedure.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received report regarding 3 concerto coils that had resistance in the middle of the catheter.The first coil (model: pv-3-8-3d, lot: b258906) was difficult to detach but was able to be deployed and detached.The second and third coil had the same resistance during delivery and could not be detached.It was noted that the devices were prepared as indicated in the instructions for use (ifu).There was no damage to the catheter but the coils were damaged.It was noted there was no issue with the instant detacher.A competitor's product was then used to complete the procedure.There was no harm or injury to the patient who was undergoing a coil embolization procedure to treat gi bleed.There were no abnormalities noted with the patient's anatomy.There was no additional medical or surgical intervention needed to prevent permanent patient impairment.The event did not extend the patient's hospitalization.
|
| |
|
Additional Manufacturer Narrative
|
|
H3.Product analysis #705219918:equipment used: vis (m-78210), 203cm ruler (m-83361) as found condition (condition of returned device): two concerto implant coils were returned for analysis within a shipping box; within their opened concerto implant coil outer cartons and within a sealed plastic bag.The (terumo) 2.8f progreat micro catheter was not returned for analysis.Visual inspection/damage location details: the (terumo) 2.8f progreat micro catheter was not returned for analysis.The minimum id (inner diameter) of the 2.8f progreat micro catheter is 0.027¿, as per an online source.Per the concerto coil ifu (instructions for use), the minimum catheter id required for use is 0.0165¿.Therefore, the progreat micro catheters are compatible for use with the concerto coils.The concerto implant coil was returned within introducer sheath.The concerto coil pushwire was found to be broken but retained by release wire at ~ 140.9c m from proximal end.The implant coil was found to be damaged within introducer sheath.The shield coil was found to be intact.The implant coil was unable to be pushed out from introducer sheath for further analysis as the implant coil was found to be stuck.All other subassemblies appeared to be normal, and no other anomalies were observed.Testing/analysis (including sem reports): the concerto coils could not be used for resistance testing with an in-house micro catheter due to their damaged condition.Conclusion: based on the device analysis and the reported information, the customers report of ¿catheter resistance¿ could not be confirmed and the root cause could not be determined.Possible causes for coil resistance in catheter are, but not limited to, failure to hydrate the concerto coil prior to use, and lack of continuous flush during delivery.H6.Coding updated based on analysis results.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported that patient's vessel tortuosity was moderate, and the coils had come out of the introducer sheath smoothly.The microcatheter used in the event was a progreat 2.8fr.No detachment attempts were made as the coils got stuck inside the microcatheter.
|
| |
|