| Model Number |
PV-2-6-3D |
| Medical Device Problem Codes |
Separation Failure (2547); Material Deformation (2976); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received report regarding 3 concerto coils that had resistance in the middle of the catheter.The first coil (model: pv-3-8-3d, lot: b258906) was difficult to detach but was able to be deployed and detached.The second and third coil had the same resistance during delivery and could not be detached.It was noted that the devices were prepared as indicated in the instructions for use (ifu).There was no damage to the catheter but the coils were damaged.It was noted there was no issue with the instant detacher.A competitor's product was then used to complete the procedure.There was no harm or injury to the patient who was undergoing a coil embolization procedure to treat gi bleed.There were no abnormalities noted with the patient's anatomy.There was no additional medical or surgical intervention needed to prevent permanent patient impairment.The event did not extend the patient's hospitalization.
|
| |
|
Additional Manufacturer Narrative
|
|
Treating physician's information was not reported.Associated with device reported in rr #: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
Additional information received reported that patient's vessel tortuosity was moderate, and the coils had come out of the introducer sheath smoothly.The microcatheter used in the event was a progreat 2.8fr.No detachment attempts were made as the coils got stuck inside the microcatheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|