| Model Number |
COBAS 8000 CORE |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable elecsys hcg+ss results for 8 patient samples on a cobas 8000 core unit.Sample 1: the initial result was 10000 miu/ml with a >test data flag.A 1:20 dilution was performed and the result was 15.4 miu/ml.The sample was retested and the result was 10000 miu/ml with a >test data flag.The repeated result was 12545 miu/ml.Sample 2: the initial result was 10000 miu/ml with a >test data flag.A 1:20 dilution was performed and the result was 53.3 miu/ml.The sample was retested and the result was 10000 miu/ml with a >test data flag.The repeated result was 48190 miu/ml.Sample 3: the initial result was 10000 miu/ml with a >test data flag.A 1:20 dilution was performed and the result was 72.2 miu/ml.The sample was retested and the result was 10000 miu/ml with a >test data flag.The repeated result was 37854 miu/ml.The sample was tested again and the result was 0.200 miu/ml with a test data flag.A 1:20 dilution was performed and the result was 30.6 miu/ml.The sample was retested and the result was 10000 miu/ml with a >test data flag.Sample 5: the initial result was 20.9 miu/ml.A 1:20 dilution was performed and the result was 25025 miu/ml.Sample 6: a 1:20 dilution was performed and the result was 166 miu/ml.The sample was repeated with a 1:20 dilution and the result was 65672 miu/ml.Sample 7: a 1:20 dilution was performed and the result was 38 miu/ml.The sample was repeated with a 1:20 dilution and the result was 35016 miu/ml.Sample 8: a 1:20 dilution was performed and the result was 27.8 miu/ml.The sample was repeated with a 1:20 dilution and the result was 22983 miu/ml.The reagent lot number is 584575.The expiration date was not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Facility name: (b)(6).The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation did not identify a product problem.The cause of the event could not be determined.A reagent related issue was excluded as the same reagent was used for the initial and repeat results.
|
| |
|
Search Alerts/Recalls
|