• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM 4MAX REPERFUSION CATHETER; NRY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM 4MAX REPERFUSION CATHETER; NRY Back to Search Results
Catalog Number 4MAXC
Medical Device Problem Codes Break (1069); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the right m2 segment of the middle cerebral artery (mca) using a penumbra system 4max reperfusion catheter (4maxc) and a neuron max 6f 088 long sheath (neuron max).During the procedure, the physician placed the neuron max in the patient.Next, while advancing the 4maxc into the neuron max, the physician experienced resistance.Therefore, the physician decided to remove the 4maxc.Upon removal, the physician noticed that the mid-shaft of the 4maxc was broken.The procedure was completed using a new 4maxc and the same neuron max.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned 4maxc confirmed a fracture on its mid-shaft.If the 4maxc is advanced at an extreme angle during use, damage such as a kink may occur.This damage likely contributed to the reported resistance during advancement.Subsequently, further manipulation of a kinked device may worsen to a fracture.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM 4MAX REPERFUSION CATHETER
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key15457681
Report Number3005168196-2022-00439
Device Sequence Number15173883
Product Code NRY
UDI-Device Identifier00814548012414
UDI-Public814548012414
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K160449
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/20/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/27/2024
Device Catalogue Number4MAXC
Device Lot NumberF109492
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2022
Initial Report FDA Received Date09/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age63 YR
Patient SexMale
Date Report Sent to FDA01/01/2005
-
-