MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER EDGE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER EDGE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES Back to Search Results
Model Number E2450H-GNSB
Medical Device Problem Code Fire (1245)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, the tip was not inside the patient when the flame occurred.The tissue was clamped with a stile clamp, the cautery (b)(4) was above the skin touching the clamp to cauterize tissue.The nurse noticed the tip began smoking and started sparking and then caught fire.The flame was about a match size.It burned in a very short time like 5 seconds, she fanned the flame out.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during procedure, the tip was not inside the patient when the match size flame occurred.Tissue was clamped with a stile clamp, the cautery (e2450hnsb) was above the skin touching the clamp to cauterize tissue.The nurse noticed the tip began smoking and started sparking and then caught fire.The device burned in a very short time like 5 seconds and was flamed out when the nurse fanned the fire.The procedure was delayed for about 1 minute so another cautery could be opened.The generator output was set to 50/50 and the long activation cycles was utilized for maximum of 10 seconds.Patient was asleep on the or table.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDGE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key15458695
Report Number1717344-2022-01149
Device Sequence Number17849380
Product Code GEI
UDI-Device Identifier10884524002248
UDI-Public10884524002248
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K962044
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE2450H-GNSB
Device Catalogue NumberE2450H-GNSB
Device Lot Number212980296R
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2022
Supplement Date Received by Manufacturer09/09/2022
Initial Report FDA Received Date09/20/2022
Supplement Report FDA Received Date09/27/2022
Date Device Manufactured10/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VLFT10GEN FT SERIES (LOT NUMBER: T5G00373DX).
Patient SexUnknown
-
-