BOSTON SCIENTIFIC CORPORATION SOLOIST; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 36703
Medical Device Problem Code Therapeutic or Diagnostic Output Failure (3023)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that insufficient roll-off occurred and the procedure was cancelled.The target lesion was located in the osteoid osteoma in the right leg tibia bone.The soloist needle was advanced through a 13 gauge 10cm biopsy needle.After positioning the soloist needle, the biopsy needle was pulled back 1.5 cm and ablation was started in the lesion.Lesion size was 7 mm.After the first ablation cycle of 15 min, no impedance was achieved.The impedance remained constant at 260 to 270.The 2nd cycle was started from 70% of the highest set power of 1st cycle, and continued for 5 min at the same power then start increasing power as 1watt/min, but impedance was not achieved.Impedance remained constant at 260 to 270.The same process was repeated 5 times and impedance was not achieved.The procedure was not completed due to this event.There were no patient complications and the patients status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLOIST
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15459331
Report Number2124215-2022-36199
Device Sequence Number6005453
Product Code GEI
UDI-Device Identifier08714729122487
UDI-Public08714729122487
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K053128
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/26/2025
Device Model Number36703
Device Catalogue Number36703
Device Lot Number0029303737
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/26/2022
Initial Report FDA Received Date09/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age16 YR
Patient SexMale
Patient Weight32 KG
-
-