MAUDE Adverse Event Report: SMITH & NEPHEW, INC. INTERTAN 10MM X 24CM 125D; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. INTERTAN 10MM X 24CM 125D; NAIL, FIXATION, BONE Back to Search Results
Model Number 71934479
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Perforation (2001)
Date of Event 02/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, following a trauma revision surgery where a lag/comp screw kit was exchanged due to early painful cutout, slight penetration into the acetabulum and a sustained femoral neck fracture on (b)(6) 2021 ((b)(4)), the patient experienced a stable collapse of the superior aspect of the femoral head with screw intrusion into the hip joint as stated within notes related to an office visit conducted in or around (b)(6) 2022 where postoperative x-rays were analyzed.The surgeon recommended either a hardware removal or a conversion to a total hip replacement system as options to treat these complications, but the patient was not interested in these options.Further details regarding the patient's final outcome are not known.This incident was noticed in a retrospective post-market clinical follow up activity (pmcf)], additional information is not known, and it is not possible to collect it, we do not have the opportunity to revisit negative feedback to retrieve further information.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
The device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, the patient experienced a stable collapse of the superior aspect of the femoral head with screw intrusion into the hip joint approximately one year post revision of the lag/comp screw (during which the acetabulum was slightly penetrated along with a femoral neck fracture).Reportedly, the patient declined the options of hardware removal or a conversion to a total hip replacement system.As of the date of this medical investigation no relevant supporting clinical information has been provided to assist with this clinical investigation as the data was reported from a retrospective post market clinical follow-up.Therefore, no clinical factors could be assessed which would have contributed to the event.With the limited information provided, the patient impact beyond the reported femoral head collapse and screw intrusion into the hip joint cannot be determined.No further medical assessment can be rendered at this time.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the instructions for use documents for intramedullary nail system revealed that provides complete guidelines of indications, contraindications, warnings and precautions and possible adverse effects that may occur preoperative, during surgery or post operative.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include traumatic injury, joint tightness, patient condition or postoperative care.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERTAN 10MM X 24CM 125D
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15459779
Report Number1020279-2022-04126
Device Sequence Number17535127
Product Code JDS
UDI-Device Identifier00885556105092
UDI-Public00885556105092
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71934479
Device Catalogue Number71934479
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2022
Supplement Date Received by Manufacturer10/21/2022
Initial Report FDA Received Date09/20/2022
Supplement Report FDA Received Date10/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-