Model Number RFG-NT-2000
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
During processing of this incident, attempts were made to obtain complete patient information.
 
Event or Problem Description
During the procedure, the generator started beeping and smoking.It was turned off and replaced to complete the procedure.There was no adverse consequence to the patient.
 
Additional Manufacturer Narrative
Based on the information provided to abbott and the investigation performed, the root cause of the reported event was isolated to thermal damage on the rf controller board.Control board.Capa 126183 is related to this complaint investigation.The issue will continue to be monitored for trends and reviewed in qdr (quality data review).The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4 LESION NT2000¿ PAIN MANAGEMENT RF GENERATOR
Common Device Name
GENERATOR, LESION, RADIOFREQUENCY
Manufacturer (Section D)
ST. JUDE MEDICAL, INC. (AF-ST. PAUL)
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC. (AF-ST. PAUL)
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key15460221
Report Number2184149-2022-00226
Device Sequence Number4274169
Product Code GXD
UDI-Device Identifier05415067022417
UDI-Public05415067022417
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K111576
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRFG-NT-2000
Device Catalogue NumberRFG-NT-2000
Device Lot Number8109802
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2022
Supplement Date Received by Manufacturer09/21/2022
Initial Report FDA Received Date09/20/2022
Supplement Report FDA Received Date09/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/20/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexFemale
-
-