BECTON DICKINSON MEDICAL (SINGAPORE) BD ECLIPSE¿ NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 305767
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd eclipse¿ needle fell off and unable to safely secure.The following information was provided by the initial reporter: one of our nurses was giving an injection using a bd eclipse needle 25g 1 ½ inch.During the injection, the safety device fell off and she was unable to safely secure the needle post-injection.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 09-sep-2022.A device history review was conducted for lot number 1204253.Our records show that this is the only instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.Additionally, three representative samples were submitted to aid in our investigation.During visual analysis and functional testing of the returned units, our engineers were unable to identify any detectable damage or other abnormalities.Unfortunately without the ability to review the reported non-conformance our quality engineers were unable to determine the root cause for this complaint.H3 other text : see h10.
 
Event or Problem Description
It was reported that the bd eclipse¿ needle fell off and unable to safely secure.The following information was provided by the initial reporter: one of our nurses was giving an injection using a bd eclipse needle 25g 1 ½ inch.During the injection, the safety device fell off and she was unable to safely secure the needle post-injection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ECLIPSE¿ NEEDLE
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15460386
Report Number8041187-2022-00550
Device Sequence Number2892007
Product Code FMI
UDI-Device Identifier30382903057673
UDI-Public(01)30382903057673
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K161170
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/31/2026
Device Model Number305767
Device Catalogue Number305767
Device Lot Number1204253
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2022
Supplement Date Received by Manufacturer09/27/2022
Initial Report FDA Received Date09/21/2022
Supplement Report FDA Received Date10/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-