• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERYAN MEDICAL LTD BIOMIMICS 3D VASCULAR STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERYAN MEDICAL LTD BIOMIMICS 3D VASCULAR STENT SYSTEM Back to Search Results
Model Number 142122-06
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Restenosis (4576)
Date of Event 03/07/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
There was no reported device malfunction and the device was not returned for analysis.A review has been performed of the manufacturing certificate of conformance associated with this lot.There were no non-conformances or deviations which could be associated with this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effect of restenosis leading to intervention is listed in the biommics 3d instructions for use and is a known patient effect of peripheral stenting procedures.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.
 
Event or Problem Description
The patient was treated as part of the mimics 3d usa post-market observational study on (b)(6) 2021.The subject was implanted with one biomimics 3d (bm3d) stent (6.0 x 80mm stent) to treat a denovo lesion of the sfa middle third in the right leg.A contralateral approach was used and the target lesion was prepared by pre-dilation of the target lesion with a standard balloon percutaneous transluminal angioplasty and atherectomy.On the (b)(6) 2022, an event of restenosis of treated vessel (target vessel) was reported from the site.It was reported as "not related" to the device and "not related" to the procedure but due to a worsening pre-existing condition.It was also reported as not target lesion related.On (b)(6) 2022 the site provided an update to the event that highlighted the event led to a percutaneous intervention and it involved a target lesion revascularisation on (b)(6) 2022 with laser atherectomy.The device remains implanted.The patient outcome is reported as "resolved/recovered".The event was reviewed on (b)(6) 2022 and considered possibly related to the device.
 
Additional Manufacturer Narrative
There was no reported device malfunction and the device was not returned for analysis.A review has been performed of the manufacturing certificate of conformance associated with this lot.There were no non-conformance's or deviations which could be associated with this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effects of restenosis leading to intervention and claudication/leg pain are listed in the bionomics 3d instructions for use and are known patient effects of peripheral stenting procedures.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.
 
Event or Problem Description
The patient was treated as part of the mimics 3d usa post-market observational study on (b)(6) 2021.The subject was implanted with one biomimics 3d (bm3d) stent (6.0 x 80mm stent) to treat a denovo lesion of the sfa middle third in the right leg.A contralateral approach was used and the target lesion was prepared by pre-dilation of the target lesion with a percutaneous transluminal angioplasty/standard balloon angioplasty and atherectomy.An event of restenosis of treated vessel (target vessel) was identified by the site on (b)(6) 2022.It was reported as "not related" to the device and "not related" to the procedure but due to a worsening pre-existing condition.It was also reported as target lesion related.The patient experienced claudication in the right leg extremity (rle) and presented for ultrasound and an angiogram was performed.The event led to a percutaneous intervention and it involved a target lesion revascularisation on (b)(6) 2022 with laser atherectomy and pta/standard balloon angioplasty on the sfa middle third.The device remains implanted.The patient outcome is reported as "resolved/recovered".The event was reviewed on 22-aug-2022 and considered possibly related to the device.
 
Additional Manufacturer Narrative
There was no reported device malfunction and the device was not returned for analysis.A review has been performed of the manufacturing certificate of conformance associated with this lot.There were no non-conformances or deviations which could be associated with this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effects of restenosis leading to intervention and claudication/leg pain are listed in the biomimics 3d instructions for use and are known patient effects of peripheral stenting procedures.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.
 
Event or Problem Description
The patient was treated as part of the mimics 3d usa post-market observational study on (b)(6) 2021.The subject was implanted with one 6.0 x 80 mm biomimics 3d (bm3d) stent to treat a denovo lesion of the superficial femoral artery (sfa) middle third in the right leg.A contralateral approach was used and the target lesion was prepared by pre-dilation of the target lesion with a percutaneous transluminal angioplasty (pta) balloon and atherectomy.An event of restenosis of treated segment (target lesion) was identified by the site on 07-mar-22.It was reported as "not related" to the device and "not related" to the procedure but due to a worsening pre-existing condition.It was also reported as target lesion related.The patient experienced claudication in the right leg extremity (rle) and presented for ultrasound and an angiogram was performed.The event led to a percutaneous intervention and it involved a target vessel/lesion revascularisation (tlr/tvr) on 01-apr-22 with laser atherectomy and pta/standard balloon angioplasty on the sfa middle third.The device remains implanted.The patient outcome is reported as "resolved/recovered".The event was reviewed on 22-aug-22 and considered possibly related to the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMIMICS 3D VASCULAR STENT SYSTEM
Common Device Name
BIOMIMICS 3D VASCULAR STENT SYSTEM
Manufacturer (Section D)
VERYAN MEDICAL LTD
block 5
parkmore east business park
galway, H91 V 0TX
EI  H91 V0TX
Manufacturer Contact
alan mcdonagh
block 5
parkmore east business park
galway, H91 V-0TX
EI   H91 V0TX
MDR Report Key15460707
Report Number3011632150-2022-00087
Device Sequence Number19950405
Product Code NIP
UDI-Device Identifier05391526850428
UDI-Public(01)05391526850428(17)220409(11)200908(10)0000051495
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P180003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/09/2022
Device Model Number142122-06
Device Catalogue Number142122-06
Device Lot Number0000051495
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2022
Supplement Date Received by Manufacturer06/16/2023
08/15/2023
Initial Report FDA Received Date09/21/2022
Supplement Report FDA Received Date07/14/2023
09/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/08/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASPIRIN.; ASPIRIN.; CLOPIDOGREL (PLAVIX).; CLOPIDOGREL (PLAVIX).; HEPARIN.; HEPARIN.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient SexFemale
Patient Weight54 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-