Model Number ZZ19KN70ZZZ006N
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a pre-use check, the customer noticed air was leaking from the product.No patient injury reported.
 
Additional Manufacturer Narrative
Visual inspection: the sample was visually inspected at a distance of 12" under normal lighting to received unit, in order to detect any damage on the unit.No discrepancies detected.Functional test: sample was tested using lt-4-2-072, the following gauge from of lt-4-2-072 was used for the test: the circuit did not pass the leak test.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
aichi pref.
minneapolis, MN 55442
MDR Report Key15461073
Report Number3012307300-2022-19910
Device Sequence Number11525777
Product Code JOH
UDI-Device Identifier15019517209983
UDI-Public15019517209983
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberZZ19KN70ZZZ006N
Device Catalogue NumberZZ19KN70ZZZ006N
Device Lot Number3881190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2022
Initial Report FDA Received Date09/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-