| Model Number |
8884752042 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Patient user of gastrostomy, states that since changing the connector there is a leak of the nutrition.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record review of the reported lot number shows evidence that the product was released according to all established procedures and qa documentation.A sample analysis could not be performed since no photo samples nor actual sample device was received for evaluation.The root cause could not be determined without a sample to evaluate.The investigation was carried out with a multifunctional team, and the findings were that all processes and controls were found to be followed according to procedure, a 100% visual inspection by production personnel according to standard work instructions.In addition, a qc sampling inspection is performed to ensure that the product has no leaks.A risk assessment was carried out to determine the observed occurrence for this failure mode.Based on the analysis, the observed occurrence of harm (o) is 0.00117% (1, improbable) which is equal to the range of the expected occurrence of harm (0) (1, improbable).This complaint shall be reopened if a sample is received.
|
| |
|