| Catalog Number |
110030776 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Unintended Movement (3026)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a revision procedure approximately 10 months postimplantation due to impingement of the glenosphere and custom glenoid.Glenosphere and vrs were removed and the tray and bearing swapped with a versadial head.Attempts have been made and additional information on the reported event is unavailable at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Procode: phx.Concomitant medical products:d10: compr vrs glen pps min tpr adr cat# 110027734 lot# 604210.Visual examination of the provided pictures and videos identified the bearing is worn secondary to impinging on the bone.However, as the product was not returned, further visual and dimensional analysis could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: multiple glenoid implant fixation screw fractures.Mild glenoid implant displacement.Suspected metal on metal implant contact.The humeral implant fit is maintained.There is apparent glenoid implant loosening and mild displacement secondary to multiple fractured fixation screws.Bone quality is osteopenic.There appears to be implant impingement.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional/corrected information.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a revision procedure approximately 10 months post implantation due to impingement of the glenosphere and custom glenoid along with the fracturing of the screws.Glenosphere and vrs were removed along with all screws.The tray and bearing were swapped with a versadial head.
|
| |
|