| Model Number |
1960-50-450 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Codes |
Pain (1994); Unspecified Musculoskeletal problem (4535); Unspecified Tissue Injury (4559); Swelling/ Edema (4577)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).(b)(4) medical records and additional information for ad 1 sep 2022, ad 8 sep 2022, ad 9 sep 2022 and ad 13 sep 2022 were reviewed by clinician.On (b)(6) 2012, the patient had a right total knee arthroplasty to address arthrosis.Depuy components were utilized, including depuy patella and depuy cement x1.((b)(4) medical records ad 13 sep 2022.Pdf part/lot).(b)(4) additional information ad (b)(6) 2022 (1) notes the patient had an aspiration with evaluation for infection, which was negative.Radiographs was noted to show a displacement of the patella button.The patient presented with pain, swelling, and dysfunction.On (b)(6) 2022, the patient had a revision right knee.During the procedure the surgeon observed a patella tendon rupture and an oblique tear of the patellar tendon.Hemosiderin stained joint fluid was noted, the patella component had been sheared from patella, the three pegs on the back of the implant were still imbedded in the patella and the remainder of the implant had displaced.All components were revised.The femoral component was found to be loose.Depuy components were implanted during this procedure, including depuy patella.
|
| |
|
Event or Problem Description
|
|
It was reported that the consumer had a replacement for left and right knee.Consumer have been in so much pain on this knee.The knee that they put it was a bad knee.Doi: (b)(6) 2012, dor: (b)(6) 2022, right knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.All available photographs were reviewed and found insufficient evidence to confirm implant loosening.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation mre was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b5, h6 health effect - clinical code.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Medical records received.(b)(6) 2022: patient felt a pop in her right knee with subsequent pain and swelling.Physician states the patient has unknown loosening of the tka components.X-ray shows migration of the patella and impingement into the patella tendon.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: corrected: d4 (primary udi number).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|