NEUWAVE MEDICAL, INC. NEUWAVE PR XT PROBE 20CM 15GA; SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES Back to Search Results
Model Number PR20XT
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when the facility was pulling devices for an ablation case it was noticed that the product box was for pr20xt but the product inside was pr15xt and the label on the box says pr15xt.There were no consequences reported.
 
Additional Manufacturer Narrative
(b)(4).Batch #: unknown.Investigation summary: the probe was returned, where is was confirmed that the barcode, tyvek, and probe inside were all for pr15xt, and the square label on the boxes were for pr20xts.There is no lot information available at this time.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 1/6/2023.A search of nonconformities (ncs) was performed for lot ml22027437.One nc was created from this complaint: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUWAVE PR XT PROBE 20CM 15GA
Common Device Name
SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES
Manufacturer (Section D)
NEUWAVE MEDICAL, INC.
3529 anderson st
madison WI 53704
Manufacturer (Section G)
NEUWAVE MEDICAL, INC.
3529 anderson st
madison 53704
Manufacturer Contact
orla o'mahony
3529 anderson st
madison, WI 53704
329348013
MDR Report Key15462636
Report Number3008769756-2022-00050
Device Sequence Number6873885
Product Code NEY
UDI-Device Identifier00853719006425
UDI-Public853719006425
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K160936
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPR20XT
Device Catalogue NumberPR20XT
Device Lot NumberML22027437
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/21/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/20/2022
Supplement Date Received by Manufacturer12/21/2022
Initial Report FDA Received Date09/22/2022
Supplement Report FDA Received Date01/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-