| Model Number |
09H39-01 |
| Medical Device Problem Code |
Smoking (1585)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer observed smoke from outlet on the side of the instrument where the reagents are located on cell-dyn emerald analyzer.The customer smelled burnt smell.The screen of analyzer blacked out.There was no visible fire or spark.There was no patient involvement and no harm to user reported.
|
| |
|
Additional Manufacturer Narrative
|
|
During the site visit by the field service engineer (fse) inspected the analyzer and removed the cpu board, part number 8701686907 and visually inspected it and verified there were no damaged components or signs of fire.Cpu board was reinstalled, and instrument is up and running without any issues.There was no fire or smoke was seen nor any damage to the instrument.There was no injury to personnel reported.A review of the service history for the cell dyn emerald analyzer revealed no additional issues were reported.A review of tracking and trending did not identify any trends for the cell dyn emerald analyzer or cpu board.Based on the available information, no systemic issue or deficiency of the cpu board or the cell dyn emerald analyzer was identified.
|
| |
|