• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT LABORATORIES CELL-DYN EMERALD SYSTEM; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT LABORATORIES CELL-DYN EMERALD SYSTEM; COUNTER, DIFFERENTIAL CELL Back to Search Results
Model Number 09H39-01
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event or Problem Description
The customer observed smoke from outlet on the side of the instrument where the reagents are located on cell-dyn emerald analyzer.The customer smelled burnt smell.The screen of analyzer blacked out.There was no visible fire or spark.There was no patient involvement and no harm to user reported.
 
Additional Manufacturer Narrative
During the site visit by the field service engineer (fse) inspected the analyzer and removed the cpu board, part number 8701686907 and visually inspected it and verified there were no damaged components or signs of fire.Cpu board was reinstalled, and instrument is up and running without any issues.There was no fire or smoke was seen nor any damage to the instrument.There was no injury to personnel reported.A review of the service history for the cell dyn emerald analyzer revealed no additional issues were reported.A review of tracking and trending did not identify any trends for the cell dyn emerald analyzer or cpu board.Based on the available information, no systemic issue or deficiency of the cpu board or the cell dyn emerald analyzer was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL-DYN EMERALD SYSTEM
Common Device Name
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
ABBOTT LABORATORIES
4551 great america parkway
santa clara CA 95054
Manufacturer (Section G)
ABBOTT LABORATORIES
4551 great america parkway
santa clara CA 95054
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key15462683
Report Number2919069-2022-00029
Device Sequence Number15173773
Product Code GKZ
UDI-Device Identifier00380740020088
UDI-Public00380740020088
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K081495
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number09H39-01
Device Catalogue Number09H39-01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/01/2022
Supplement Date Received by Manufacturer09/26/2022
Initial Report FDA Received Date09/22/2022
Supplement Report FDA Received Date10/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/18/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-